3-DUS和TUI超声技术在胎儿畸形筛查中的应用价值
DOI:
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:


Application of 3-DUS and TUI ultrasonography to fetal malformation screening
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    探讨表面三维成像技术(3-DUS)与断层超声显像技术(TUI)在筛查胎儿畸形中的应用价值。方法选取1 896 例于该院行三维超声检查的孕妇为研究对象,采用3-DUS 技术与TUI 技术检查胎儿畸形情况。结果胎儿畸形发生率为3.22%,体表畸形发生率为65.57%,体内畸形发生率为34.43%;3-DUS 技术体表畸形阳性检出率为97.50%,体内畸形检出率为19.05%,TUI 技术体表畸形阳性检出率为42.50%,体内畸形检出率为100%;3-DUS 技术诊断胎儿畸形敏感性、特异性分别为70.49%和99.46%,TUI 技术诊断胎儿畸形敏感性、特异性分别为62.30%和99.35%,两者比较,差异无统计学意义(p >0.05),3-DUS 与TUI 技术联合诊断胎儿畸形敏感性、特异性分别为100%和99.89%,其敏感性高于单纯3-DUS 及单纯TUI 诊断(p <0.05),特异性高于单纯3-DUS 及单纯TUI诊断,但差异无统计学意义(p >0.05)。结论3-DUS 技术诊断胎儿体表畸形检出率高,TUI 技术诊断胎儿体内畸形检出率高,两者联合在胎儿畸形筛查中具有较高应用价值。

    Abstract:

    To investigate the application of 3-DUS and TUI ultrasonography in fetal malformation screening. Methods Totally 1,896 pregnant women in our hospital were selected as the research objects. 3-DUS technique and TUI technique were used to examine fetal malformations. Results The rate of fetal malformations was 3.22%, in which surface malformations accounted for 65.57% and internal malformations accounted for 34.43%. Using 3-DUS the positive detection rate of surface deformities was 97.50%, that of internal malformations was 19.05%. Using TUI the positive detection rate of surface malformations was 42.50%, that of internal malformations was 100%. The sensitivity and specificity of 3-DUS in the diagnosis of fetal malformations were 70.49% and 99.46% respectively, those of TUI in the diagnosis of fetal malformations were 62.30% and 99.35% respectively, there were no significant differences between the two techniques (p > 0.05). The sensitivity and specificity of 3-DUS combined with TUI in the diagnosis of fetal abnormalities were 100.00% and 99.89%, the sensitivity was significantly higher than that of pure 3-DUS diagnosis and pure TUI diagnosis (p < 0.05), the specificity was also higher than that of pure 3-DUS diagnosis and pure TUI diagnosis but without significant differences (p > 0.05). Conclusions 3-DUS technique has a high detection rate of fetal surface malformations, while TUI technique has a high detection rate of fetal internal malformations. The combination of the two techniques has high application value in screening of fetal malformations.

    参考文献
    相似文献
    引证文献
引用本文

丁媛,远秀芹,刘兰兰.3-DUS和TUI超声技术在胎儿畸形筛查中的应用价值[J].中国现代医学杂志,2017,(16):126-128

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2017-02-08
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2017-08-10
  • 出版日期: